Dew point meters are instruments dedicated to measuring the temperature at which the water vapor in the air begins to condense — a fundamental parameter for controlling climatic conditions before and during the application of coatings, for the qualification of compressed air, for monitoring industrial gases and for air-conditioning applications. Knowing the dew point is essential in order to avoid unwanted condensation on cold surfaces, to know whether the air is too humid or too dry for a specific process, and to guarantee compressed-air quality for critical applications.

Would you like help choosing the right product ?
Measuring technologies
The technologies used are: indirect measurement from T and RH (combined T+RH instruments calculate the dew point using psychrometric formulas — typical accuracy ±2°C); the chilled-mirror hygrometer (primary reference: a mirror cooled by a Peltier or thermoelectric cell is brought below the dew point until condensation forms and is detected optically, and the mirror temperature at equilibrium is the dew point — accuracy ±0.1-0.2°C); polymer capacitive sensors (for low-humidity measurements in compressed air and industrial gases — typical range -80°C / +20°C); aluminum oxide Al2O3 sensors (for dry process gases, range down to -100°C); silicon oxide sensors (for high-end applications).
Climatic applications
For industrial painting applications, verification of the dew point is prescribed by ISO 8502-4: the temperature of the metal substrate must be at least 3°C above the dew point of the ambient air in order to ensure the absence of condensation during application of the coating. Even minimal condensation compromises adhesion of the primer and causes serious defects (blistering, detachment). DPM (Dew Point Meter) instruments for painting measure air temperature, relative humidity and substrate temperature simultaneously and calculate the T-DP delta, providing alarms if the conditions are not acceptable.
Compressed air and industrial gases
For compressed-air applications the dew point is the key quality parameter: process air must be dry in order to avoid condensation in pipework and pneumatic actuators, ice formation at expansions, and contamination of sensitive processes. ISO 8573-1 classifies air quality according to the dew point under pressure (for example class 3: dew point ≤ -20°C). Dedicated instruments measure the dew point under pressure (calculating the atmospheric dew point for comparison with ISO 8573-1) and are used in the qualification of compressors, dryers and compressed-air systems.
Portable instruments
Portable dew point meters integrate dedicated sensors and displays for calculating the dew point, with the following functions: simultaneous calculation of T, RH, DP, T-DP and wet-bulb T; programmable alarm thresholds for painting conditions; ISO 8502-4 mode with automatic verification of the 3°C safety delta; reading memory; USB/Bluetooth data output for archiving; integrated or separate surface probes for measuring substrate temperature. Rugged models for site use and high-precision laboratory models are both available.
Reference standards
The main standards are: ISO 8502-4 for industrial painting (verification of climatic conditions); ISO 8573-1 for compressed-air quality; ISO/CD 21501 for air quality control in the pharmaceutical sector; OIML R121 for hygrometers. Calibration of chilled-mirror instruments is carried out at high-level metrological laboratories (ISO 17025 accredited for dew point). Capacitive instruments are calibrated with humidity generators or by comparison with a reference dew point meter. The typical calibration interval is annual.
Selection criteria
Selecting a dew point meter depends on the application: industrial painting → DPM instruments with a substrate temperature probe; compressed air → instruments for measurement under pressure compliant with ISO 8573; industrial gases at low humidity → instruments with an Al2O3 sensor for extended ranges; laboratory applications → chilled-mirror instruments as a primary reference. Portability, robustness, battery life and the availability of an ISO 17025 certificate are essential factors in the choice.